2.1.3 Measurements

Measurement specifications apply to single chassis AC sources. See notes for other models and
configurations. Frequency, VA, power factor and crest factor only apply in AC or AC+DC modes.

Parameter

Range

Accuracy (

% FS)

Resolution

Frequency

1

16.00-1000.0 Hz

0.1 %

0.01 Hz to 81.91 Hz
0.1 Hz to 819.1 Hz
1 Hz > 819.1 Hz

Voltage

0 - 400 Volts

0.1 %

0.01 Volt

Current

0

– 15 Amps

0.5 %

0.001 Amp

Peak Current

0 - 60 Amps

0.5 %

0.001 Amp

Crest Factor

1.00

–10.00

1.5 %

0.01

VA Power

0 - 4 KVA

0.5 %

1 VA

Real Power

0 - 4 KW

0.5 %

1 W

Power Factor

0.00 - 1.00

1 %

0.01

Note: Accuracy specifications are valid above 100 counts. For multi-chassis configurations, Current
and Power range and accuracy specifications are times the number of chassis.

Note: Frequency measurement specification valid for output > 20 Vrms.

Note: Crest Factor accuracy applies for Irms > 50% of max.

Note: Power Factor accuracy applies for PF > 0.5 and VA > 50% of max.

2.1.4 Harmonic Measurements

Harmonic measurement specifications apply to 751iX and 1501iX model AC sources only.

Parameter

Range

Accuracy (

% FS)

Resolution

Frequency fundamental

16.00 - 81.91 Hz
82.0 - 819.1 Hz
> 819.1 Hz

0.1%

0.01 Hz
0.1 Hz
1 Hz

Frequency harmonics

16.00 Hz

– 48 kHz

0.5%

0.1 Hz

Voltage

Fundamental 0 - 400 Volts

0.5%

0.01V

Harmonic 2 - 50

0.5% + 0.5%/kHz

0.01V

Current

Fundamental 0 - 15 Amps

0.5%

0.01A

Harmonic 2 - 50

0.5% + 0.5%/kHz

0.01A

Note: For multi-chassis configurations, current accuracy specifications are times the number of
chassis.

1

Frequency measurement specifications valid with output voltage of 20Vrms or higher. If output relay is open, frequency

measurement will return 0.0 Hz.