2.11 WHM Option - Supplemental Specification
The following measurment accuracy specifications apply to the Watt Hour meter mode of operation:
Parameter
Specification
Watt-Hour
Range: 0
– 999,999.9 WH
Resolution: 0.1 WH
Accuracy:
0.5% R +10 WH
Etime
Range:
0:00:00 to 9999:59:59
Resolution:
1 sec
Accuracy:
0.025 %
2.12 Supplemental Specifications
Supplemental specifications are provided for reference only and are not guaranteed. Data is based on
typical performance of a Compact i/iX series power source but not verified on each unit produced as
part of AMETEK Programmable Power acceptance test.
Results on individual units may vary from the data provided here.
2.12.1 Output
Output Parameter
Specification
Frequency response:
Max. Voltage slew:
4V/us, measured step response into resistive load, 10% to 90 % full-
scale voltage.
Load Transient response
No load to full load:
Voltage recovers to within 2% in less than 2 ms
Full load to no load:
Voltage recovers to within 2% in less than 2 ms
Load Regulation response in
ALC mode
< 300 ms
Load Regulation:
(ALC mode OFF)
< 3 % FS
1
1
FS (Full Scale) refers to highest available range, e.g. 300Vac in AC mode, 400Vdc in DC mode.