10.6.3 Abnormal Test

From the MIL704 main menu Figure 10-2) scroll to the ABNORMAL AC entry using the up and down
cursor keys. Press the ENTER key to select the ABNORMAL screens. The screen will appear as
shown in Figure 10-5.

Figure 10-5: Abnormal Screens

The ABNORMAL SCREEN has the following tests:

1. OVER VOLTAGE

2. UNDER VOLTAGE

3. OVER FREQUENCY

4. UNDER FREQUENCY

The above test can be selected by scrolling to the highlighted selection using the up and down key and
the ENTER key to start the selected test.

OVER VOLTAGE

This test will change the output voltage for the selected phase in the following sequence:

For 400Hz and VFREQ:

180V for 50msec.

The voltage gradually decays with time to 125 volt by the following equation:
V = 124.6 + 2.77/t. For 0.05

t

6.925

Stay at 125V for 93 seconds before returning to 115V.

For 60Hz only:

180V for 3.34msec

The Voltage gradually decays with time to 122 volt by the following equation:
V = 121.7 + 0.583/t. For 0.00334

t

1.947

Stay at 122V for 8 seconds before returning to 115V.

Note: Prior to the test, a voltage range change may take place if the power source is set for the low
voltage range. This will cause the EUT to loose power momentarily. If this is not acceptable, the power
source must be left in high range at all times.

The

key (backspace) will terminate the test at any time.