beautypg.com

AMETEK Compact i/iX Series User Manual

Page 152

background image

User Manual

Compact i/iX Series

AMETEK Programmable Power

California Instruments

152

Number

Message String

Cause

Remedy

command to clear waveform memory.

-314

"Save/recall memory
lost"

User setup register
contents lost

Store setup in same register again.

-315

"Configuration memory
lost"

Hardware configuration
settings lost.

Contact CI service department at

[email protected]

to

obtain instructions on restoring
configuration data.

-330

"Self-test failed"

Internal error

Contact CI service department at

[email protected]

-350

"Queue overflow"

Message queue full.

Too many messages. Read status using
SYST:ERR query until 0, "No Error" is
received indicating queue empty.

-400

"Query error"

Unable to complete query.

Check programming manual for correct
query format and parameters

-410

"Query INTERRUPTED"

Query issued but
response not read.

Check application program for correct
flow. Response must be read after each
query to avoid this error.

-420

"Query
UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

-430

"Query DEADLOCKED"

Query cannot be
completed

Check application program for multiple
queries

-440

"Query
UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

0

"No error"

No errors in queue

2

" Non-volatile RAM
CONFIG section
checksum failed"

Controller failure during
Self-test.

Contact CI service department at

[email protected]

3

" Non-volatile RAM CAL
section checksum failed"

Controller failure during
Self-test.

Contact CI service department at

[email protected]

4

" Non-volatile RAM
WAVEFORM section
checksum failed"

Controller failure during
Self-test.

Contact CI service department at

[email protected]

10

"Ram self test

Controller failure during
Self-test.

Contact CI service department at

[email protected]

40

"Voltage self test error,
output 1

No. 1/A amplifier in
Master source has no
output during Self-test.

Contact CI service department at

[email protected]

41

"Voltage self test error,
output 2

No. 2/B amplifier in
Master source has no
output during Self-test.
Three phase models only.

Contact CI service department at

[email protected]

42

"Voltage self test error,
output 3

No. 3/C amplifier in
Master source has no
output during Self-test
Three phase models only.

Contact CI service department at

[email protected]

43

"Current self test error,
output 1

No. 1/A amplifier in Aux.
Source has no output
during Self-test.

Contact CI service department at

[email protected]

44

"Current self test error,
output 2

No. 2/B amplifier in Aux.
Source has no output
during Self-test. Three
phase models only.

Contact CI service department at

[email protected]