FREQUENCY TRANSIENTS (A(CF) only)

Test 1

Seq. No.

Volt/Frequency

Time

1

115/400

5 minutes

2

115/350

5 sec.

3

115/320

0.2 sec.

4

0/320

0.2 sec.

5

115V/400

10 sec.

Test 2

Seq. No.

Volt/Frequency

Time

1

115/400

5 minutes

2

115/480

0.2 sec.

3

115/440

5 sec.

4

0/440

0.2 sec.

5

115V/400

10 sec.

FREQUENCY TRANSIENTS (Group 1 only)

Seq. No.

Frequency

Time

1

400

5 minutes

2

480

5 sec.

3

400

10 sec.

4

320

5 sec.

5

400

10 sec.

Table 9-14: Abnormal Frequency Transient

This test will set the voltage at 115V and will remain at this voltage through out the test except for the
A(CF) category. The test will cycle the frequency three times as shown in Table 9-14. Each repeat will
start from sequence 2. Test1 and test2 for the A(CF) category are done in succession as a single test.