4.1.3 Status Indicator Lights
Five yellow LED status indicators are located on the left hand side of the keyboard/display panel.
These LED’s correspond to the following conditions:
Hi Range
The Hi Range LED is on when the high voltage output range has
been selected.
Overtemp
The Overtemp LED indicates an overheating problem inside the unit.
This is an abnormal condition, which will cause the unit to shut off.
Check the air openings to make sure they are not blocked.
Overcurrent
The Overcurrent LED indicates an output current that exceeds the
programmed setting. This condition can be controlled by setting the
current limit value in the PROGRAM menu. Removing the load using
the OUTPUT ON/OFF button will recover from an overload condition
when in CV mode.
Remote
The Remote LED indicates that the unit is in remote control mode. If
the IEEE-488 interface is used, this indicator will be lit whenever the
ATN line (Attention) line for the GPIB address set is asserted by the
IEEE controller. If the USB or LAN interface is used, the REMOTE
state can be enabled by the controller using the SYST:REM
command. Any time the Remote LED is lit, the front panel of the
power source is disabled. Note: The BACK button doubles as a
GOTO LOCAL button (LOCAL) while the unit is in remote state. This
allows the user to regain control of the front panel. The LOCAL button
can be disabled by sending a Local Lockout bus command. This
prevents unauthorized changes of settings in ATE applications.
Output
The Output LED indicates the output relay status. If the LED is off,
the output relay is open. If the LED is on, the output relay is closed.