Number

Message String

Cause

Remedy

-200

"Execution error"

Command could not be
executed

Command may be inconsistent with
mode of operation.

-201

"Invalid while in local"

Command issued but unit
is not in remote state

Put instrument in remote state before
issuing GPIB commands.

-203

"Command protected"

Command is locked out

Some commands are supported by the
unit but are locked out for protection of
settings and are not user accessible.

-210

"Trigger error"

Problem with trigger
system.

Unit could not generate trigger for
transient execution or measurement.

-211

"Trigger ignored"

Trigger request has been
ignored.

Trigger setup incorrect or unit was not
armed when trigger was received. Check
transient system or measurement trigger
system settings.

-213

"Init ignored"

Initialization request has
been ignored

Unit was told to go to armed state but
was unable to do so. Could be caused
by incorrect transient system or
measurement acquisition setup.

-220

"Parameter error"

Parameter not allowed.

Incorrect parameter or parameter value.
Check programming manual for
allowable parameters

-221

"Setting conflict"

Transient programmed
with more than 1 mode.

Check other settings. E.g. Redefine
transient mode.
As result of *TST? execution, indicates
ALC mode is off or waveform not set to
Sine.

-222

"Data out of range"

Parameter data outside of
allowable range.

Check programming manual for
allowable parameter values

-223

"Too much data"

More data received than
expected

Check programming manual for number
of parameters or data block size

-224

"Illegal parameter value"

Parameter value is not
supported

Check programming manual for correct
parameters

-226

"Lists not same length"

One or more transient
lists programmed have
different length.

All lists must be of same length or
transient cannot be compiled and
executed.

-241

"Hardware missing"

N/A

N/A

-254

"Media full"

No storage space left to
save settings or data.

Delete other settings or data to make
room.

-255

“Directory full”

Too many waveform
directory entries

Delete one or more waveforms from
waveform memory to make room.

-256

“File name not found”

Waveform requested not
in directory

Check waveform directory for waveform
names present.

-257

“File name error”

Incorrect filename

Too many or non-ASCII characters used
in waveform file definition.

-283

“Illegal variable name”

Variable name illegal.

Use ASCII characters only

-300

"Device specific error"

Hardware related error

Check hardware for proper operation.

-311

"Memory error"

Waveform memory
checksum error.

May be the result of incomplete user-
defined waveform download. Check
interface and try downloading waveform
again. Successful download may clear
this error condition.

Alternatively, use TRAC:DEL:ALL