10.7.4 Emergency Test

From the MIL704 DC MENU scroll to the EMERGENCY DC entry using the up and down cursor keys
(Figure 10-6). Press the ENTER key to start the EMERGENCY TEST.

Figure 10-6: Emergency Test

VOLTAGE

This test will change the output voltage for the selected phase in the following sequence:

1. 28V system:

18V for 1 minute.

29V for 1 minute.

28V for 1 minute.

2. 270V system:

250V for 1 minute.

280V for 1 minute.

270V for 1 minute.

The

key (backspace) will terminate the test at any time.