ADLINK ND-6067 User Manual
Page 164
162
6.3.13 Offset Calibration
@Description
To correct the offset errors of AD converter by using the offset
calibration.
@Syntax
$(Addr)1
$
Command leading code
(Addr) Address
ID
1 Offset
calibration.
@Response
!(Addr)
or
?(Addr)
!
Command is valid.
?
Command is invalid.
(Addr) Address
ID.
@Example
User command:
$061
Response:
!06
To perform the offset calibration for analog input module at
specified address 06 (Hex).
Note: To perform the calibration, a proper input signal should
be connected to the analog input module. Different
input range have different input voltage, detail refer
chapter 5 “Calibration” .
(6013, 6017, 6018)