6.3.3 Span Calibration to each Channel
@Description
To correct the gain errors of AD converter by using the span
calibration.
@Syntax
$(Addr)0(Channel No)
$
Command leading code (1 character)
(Addr)
Address ID (2 character)
0
Span calibration (1 character)
(Channel No)
Channel for Calibration (1 character) 0~2
@Response
!(Addr)
or
?(Addr)
!
Command is valid.
?
Command is invalid.
(Addr) Address
ID.
@Example
User command:
$0601
Response:
!06
To perform the span calibration for analog input module
ND-6013 channel 1, address ID is 06H.
Note:
To perform the calibration, a proper input signal should
be connected to the analog input module. Different
input range have different input voltage, detail refer
chapter 9 “Calibration” .
(6013