ADLINK ND-6067 User Manual
Page 163
161
6.3.12 Span Calibration
@Description
To correct the gain errors of AD converter by using the span
calibration.
@Syntax
$(Addr)0
$
Command leading code (1 character)
(Addr)
Address ID (2 character)
0
Span calibration (1 character)
@Response
!(Addr)
or
?(Addr)
!
Command is valid.
?
Command is invalid.
(Addr) Address
ID.
@Example
User command:
$060
Response:
!06
To perform the span calibration for analog input module,
address ID is 06H.
Note:
To perform the calibration, a proper input signal should
be connected to the analog input module. Different
input range have different input voltage, detail refer
chapter 5 “Calibration”.
(6013, 6017, 6018)