To change the counting rate:
The following example shows how lo change the couriling rale from 0,01 Lo 0.9 I,
Key operation
Display example
Description
SEl'
□
Gl: COUNT RATE
1. OOE-Zm*
Press [SET] first to start the ilems selection sequence
and select Gl: COUNT RATE froni among the
configuration items using [ ► ] and [
a
. ] keys. Then
press [SET] again to display the current counting rate
U.Ü0E-2m-^Ü.Ol m^).
►
□
Gl: COUNT RATE
1
OOE-Zm'
Pressing [ ► ], the cursor appears. Then press [ ► ] as
many times as necessary lo Enove the cursor to of
measuring unit.
▲
□
Gl: COUNT RATE
1. OOE-Zl
fl. OOE-il
Select ‘‘1” as the measuring unit by pressing [
a
j.
(Note)
Then move the cursor to (he desired digit by pressing
[ ► J and change the value. In this example repeat this
process until the display shows “9.()0l\-11,”
SET
□
G1: COUNT RATE
9. DDE-1 ]
Pressing [SET], The cursor disappears and the new
counting rate display flickers, Press [SET] again to
store the new counting rate.
SET
Gl: COUNT RATE
Pressing [SUfj, the system returns to the items
selection sequence.
Note: The unit changes as shown below by pressing [
^
],
] —► ml----- ►
i h h \ )
-- ► (gal)-----
In case of the converter for special specifications (refer to 7.3.3 '^Converter for
Special specifications"), *'bbl" can be selected.
The unit in paremhcscs (bbl), (gal) is shown only when the meter siJte is selected in
inches.
-