beautypg.com

Toshiba – Toshiba LF434 User Manual

Page 64

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

background image

TOSHIBA

6 F . 8 A 0 7 7 0

To change the span (range):

The foJlowing example shows how to change ihc span of Range j from

2 . 1 )

m/s to 100 J/min.

Key operation

Display example

Description

SHT

C2: RANGE 1

QZ. 000 m/s

Press [SI2T] first to start the items selection sequence

and select C2: RA

j

NGE 1 from among the

cojifigtiration items using [ ► J and [

^

] keys. Then

press [SRT] again to display the current span of Range

1 (2.0 tn/s in this example).

CZ: RANGE 1

OZ. 000 i/s

Pressing [ ► ], the cursor appears, Then press [ ► ]

as many times as necessary to move the cursor to the

position for the measuring unit.

C2: RANGE 1

3. 93E+0 1/s

2. 36E+2 l/win

Select "‘1’’as the first unit of the measuring unit by
pressing 1 ] os many times as necessary. (Notcl)

Similarly, pressing | ► ] lo move the cursor to the

second unit (time unit), select “min.” (Note 2)

(The displayed span automatically changes in

accordance with the newlv selected unit.)

C2t RANGE 1

2. 3BEf2 l/min

l^esfi [ ► ] as many times as necessary to move the

cursor to the digit of span to be changed.

C2: RANGE 1

T36E+2 1/jnin

ийЙЕ+2 l/nin

Change the value by pressing [ -

a

. ], Then move the

cursor to another digit by pressing [ ► j and change Ihc

value. In this example repeat this process until the

display shows “1.{K)E+2”(=100) l/m.

SFT

C2: RANGE 1

1. OOE+2 l/min

?ressing (SETJ, the cursor disappears and the changed

display ñickers. Press [SET] again to store the changed

span and unit.

SFT

C2: RANGE 1

Pressing [SET], the system returns to the items selec­

tion sequence.

N'utes 1. The first

iniil

of the measuring unit changes as shown helow;

I---- ►

—► I---------► ml——► (bhl>

I------

- (ft)^<---------- (gal)

Units in parenlheses (bbl, ft anti gal) are shown only when Ihe meter size is sciected in

inches,

2. 'The second unit of the measuring unit changes as shown below:

/s --------- ► /min ----------►

/ U

------- ► /d—I

L

However, the follow-ing first and second unit combinations cannot be selected: nx'min,
паФ, n\''d, ft/'miti, ft/h, fi'd.

- 63 -