15. Specifications
rhe flowmcier specifications and iht type specificiilion code used when ordering the
flowmeter arc described in this chapter
15.1 Flowmeter Specifications
■ Overall Specifications
Measurement range in terms of flow velocity:
0-0.3 m/s to 0-10 m/s (0-0.1 m/s to 0-0.3 m/s range is avaibbie optionally)
System accuracy:
See the following table,
Table 15.1 System accuracy
Flow rate as a
percentage
of
range
Accuracy
11.1-1.0 m/s OJ-Ulm/s
L O - I O m / s
0 U)
2 0 %
±0.1%
i S
20 to 100%
±0.5% of rate
0 to
5 0 %
±0.25% of FS
50 to 100%
±0.5% of rate
Note: The accuracy above is measured under standard operating conditions at
Toiihiba's cfilibration facility.
Dimensions and Mass: Sec Chapter 16» ‘‘Outline Dimensions.”
- U 4 -