10.8 Remote Selection of Fixed Value Output
A user-specificd current output and pulse output can be selected witli a D1 sigiiaL
Proceed as follows to use this function:
Fixed-vulne setting;
Set tlie fixed-value for current output and for pulse output. See S.2.15,
“Hxed-Value Output/" Set the fixed-value output enable/disable status to "OFIV’
If the pulse output is not used, fixed-value setting for pulse output is not needed.
DI function setting
Set 131 to use as a fixed-value output control signal. Sec 8.2,9, “Digital FO
Control signal input conditions:
Control signal input level
4 -20 mA and pulse output
L level
Outputs the measure J value.
II level
Outputs ihe fixed-value.