7 | Maintenance

BÜCHI Labortechnik AG

92/110

PrepChrom C­700

 NOTICE

Risk of flow cell window damage.

Touching the flow cell windows with bare hands can damage the surface of the
cell window.

u

Wear protective gloves.

u

Remove both flow cell windows (4).

u

Remove all PTFE seals (5).

u

Reassemble the flow cell in reverse order to the disassembly.

u

Depending on the required thickness of the liquid layer, the installation of
the flow cell windows can be changed.

u

The optical path can be increased by using the installation positions (b) or
(c).

• Default installation position is (a).

Follow­on tasks

u

Check the cell for any impurities (dust, hairs) using a magnifier.

• Any impurity within the cell will increase the signal noise.

u

Install the flow cell according the following task:

7.2.2 Flow cell, cleaning [

}

page 88]

7.3.4 UV Detector lamp, replacement

7.3.4.1 Detector lamp, removal

Preliminary tasks

u

Purge the system with a suitable solvent (isopropanol) and with air.

u

Air purge the system.

• 6.9 Air purge [

}

page 85]