6 | Operation

BÜCHI Labortechnik AG

76/110

PrepChrom C­700

u

Select the tab Settings.

u

Select a column from the drop down menu (1).

u

Select the equilibration mode (2).

• None
• Equilibration
• Equil. + dry load
• 6.5 Equilibration and injection modes [

}

page 81]

u

Select the injection mode (3).

• Auto
• Valve injection
• Wait for injection
• Valve injection + wait
• 6.5 Equilibration and injection modes [

}

page 81]

 CAUTION

Risk of eye damage or burns due to a bursting column.

A wrong pressure setting can cause a column to burst.

u

Do not set pressure higher than the allowed maximum pressure of the col­
umn.

u

Adjust the maximum pressure (4) if necessary.

u

Select between the stop modes (5).

4.4.6 Stop modes [

}

page 32]

• Pause
• Standby
• Continue (ICM)
• Stop

u

Select an elution unit (6).