5 | Installation
BÜCHI Labortechnik AG
66/110
PrepChrom C700
5.7.13 Automatic gradient optimization, activation/deactivation
u
Press Setup > User Profile on the main menu.
u
Open Edit Method > Settings Panel > Automatic Gradient Optimiza
tion.
u
Select Show or Hide (1).
u
Press Ok.
5.7.14 Switching valve activation/deactivation
Note
The installation of the switching valve needs to be done by a BUCHI customer
service representative.
u
Press Setup > System configuration on the main menu and select the
tab Valves.
u
Click into the field Column Switch (1) and select the switching mode (2):
• SWITCH BY METHOD = flow through column 1 or 2 can be set in the
method (proceed with step 2).
• SWITCH TO COLUMN 1 = flow through column 1
• SWITCH TO COLUMN 2 = flow through column 2