5 | Installation

BÜCHI Labortechnik AG

66/110

PrepChrom C­700

5.7.13 Automatic gradient optimization, activation/deactivation

u

Press Setup > User Profile on the main menu.

u

Open Edit Method > Settings Panel > Automatic Gradient Optimiza­
tion
.

u

Select Show or Hide (1).

u

Press Ok.

5.7.14 Switching valve activation/deactivation

Note

The installation of the switching valve needs to be done by a BUCHI customer
service representative.

u

Press Setup > System configuration on the main menu and select the
tab Valves.

u

Click into the field Column Switch (1) and select the switching mode (2):

SWITCH BY METHOD = flow through column 1 or 2 can be set in the

method (proceed with step 2).

SWITCH TO COLUMN 1 = flow through column 1
SWITCH TO COLUMN 2 = flow through column 2