3 valve module, 4 control unit, 5 pump module – BUCHI PrepChrom C-700 User Manual
Page 16: 6 fraction collector, 7 electrical system
3 | Technical data
BÜCHI Labortechnik AG
16/110
PrepChrom C700
Description
Value
Maximum flow rate
500 mL/min
Time constant
1 sec.
3.2.3 Valve module
Description
Value
Loop volume
2 mL
3.2.4 Control unit
Description
Value
Screen resolution
10.1" touch screen
Screen resolution
1024×600 pixel
Operating system
Windows 7 Ultimate 32 bit
Harddisk memory
320 GB
Working memory (RAM)
4 GB
Processor
Atom D2550, Dual core, 1.86 GHz
USB
2.0
3.2.5 Pump module
Description
Value
Pressure
up to 100 bar
Flow rate range
• Flow rate accuracy 5–10 mL/min
• Flow rate repeatability 5–10 mL/min
5–250 mL/min
>4.5 %
>0.5 %
Flow rate accuracy 10–20 mL/min
<4.5 %
Flow rate accuracy 20–200 mL/min
<3.5 %
Flow rate accuracy 200–250 mL/min
<4.0 %
Flow rate repeatability 10–250 mL
<0.5 %
Gradient step accuracy 80 mL/min, 80 bar
<3.5 %
3.2.6 Fraction collector
Description
Value
Collection area (WЧD)
200Ч420 mm
3.2.7 Electrical system
Description
Value
Power consuption
350 W
Input voltage
100–120 V/220–240 V ±10 %
Frequency
50–60 Hz
Fuse (220–240 V)
TT2.5A L 250V/
TT1.25A L 250V
Fuse (100–120 V)
TT5A L 250V/TT2.5A L 250V
Overvoltage category
II