7 | Maintenance

BÜCHI Labortechnik AG

90/110

PrepChrom C­700

7.2.4 Piston cleaning disc, cleaning

 NOTICE

Risk of electrical damage.

Connecting the piston cleaning discs can cause electrical damage due to miss­
ing tubes.

u

Do not remove the plugs of the piston cleaning disc in­ and outlet.

u

The piston cleaning discs need to be installed by a BUCHI customer ser­
vice representative.

ð 8.1 Customer service [

}

page 97]

7.3 Replacement

 CAUTION

Risk of chemical burns due to high pressure in the system.

Liquid can escape from fittings or connections and minor injuries can result.

u

Wear protective goggles.

u

Wear protective gloves.

u

Wear protective clothes.

7.3.1 UV Detector cover, removal

 WARNING

Risk of electric shock.

Serious injuries can result.

u

Switch off the system.

u

Disconnect all 5 power supply cables.

u

Wait until the detector lamp is cooled down (approx. 15 minutes).

u

6.2 Switch off the system [

}

page 71].

u

Disconnect all 5 power supply cables.

u

Disconnect the RS232 cable from the UV Detector module.

u

Let the lamp cool down 15 minutes.

u

Remove the tubes from the UV Detector module.

u

Remove 2 hex screws (1), washers, spacers and the left column hold­
er (2).

u

Remove the UV Detector module and place it on a suitable place.