Hold the flow cell (2) with one hand while lifting the lever (1) further up.

u

Remove the flow cell.

u

Flush the cell with pure solvent and rinse extensively.

u

Reassemble the UV Detector module in reverse order to disassembly.

Follow­on tasks

u

Purge the system and verify there is no baseline drift.

7.2.3 Check valve, cleaning

Preliminary tasks

u

Purge the system with a suitable solvent (isopropanol).

u

Air purge the system.

6.9 Air purge [

}

page 85]

u

7.3.5.1 Check valve, removal [

}

page 94]

Tasks

NOTE

Always clean all 4 check valves.

u

Clean the check valves in a ultrasonic bath with methanol or acetone for
5 minutes.

Follow­on tasks

u

7.3.5.2 Check valve, installation [

}

page 95]