3 check valve, cleaning – BUCHI PrepChrom C-700 User Manual
Page 89

BÜCHI Labortechnik AG
Maintenance | 7
PrepChrom C700
89/110
u
Hold the flow cell (2) with one hand while lifting the lever (1) further up.
u
Remove the flow cell.
u
Flush the cell with pure solvent and rinse extensively.
u
Reassemble the UV Detector module in reverse order to disassembly.
Followon tasks
u
Purge the system and verify there is no baseline drift.
7.2.3 Check valve, cleaning
Preliminary tasks
u
Purge the system with a suitable solvent (isopropanol).
u
Air purge the system.
u
7.3.5.1 Check valve, removal [
Tasks
NOTE
Always clean all 4 check valves.
u
Clean the check valves in a ultrasonic bath with methanol or acetone for
5 minutes.
Followon tasks
u
7.3.5.2 Check valve, installation [