4.4.6 Stop modes

Corresponds to the way the system behaves at the end of the method, namely
the time of the last elution step and the possible actions done. It does not con­
cern the manual stop by clicking on the stop button in the tool bar (by default:
Pause).

Pause

When the method is finished, the message Method ended: Do you wish to
continue?
 is displayed on the screen and the pump immediately stops. Answer
No stops the method and answer Yes restarts the pumping with the last operat­
ing parameters.

Standby

The pump keeps running until the user selects pause or stop.

Continue (ICM)

At the end of the method, the pump automatically continues. The pump keeps
on with the last operating parameters during a time fixed to 10 minutes (config­
urable in the User Profile window) in order to check the signal activity (peak):

• If none activity is detected during 10 minutes, the method stops.

• If an activity higher or equal to the lower activity detected during the method

execution happens, it is managed according to the method parameters and
the ICM Stop mode restarts for 10 minutes again.

Stop

When the last elution step is reached, the method stops.

When the stop mode is chosen Continue or Stop, another automatic action be­

comes available at the end of the method (

 enabled, 

 disabled).