6.5 Equilibration and injection modes

u

Choose an equilibration mode  to condition the column before injection.

Mode

Action

None

No column conditioning before injec­
tion

Equilibration

Column conditioning before injection;
time designation

Equil. + Dry load

Column conditioning before injection;
no timing; no possibility for loop injec­
tion at the 6­way electrical valve

u

Choose injection mode 

Mode

Action

Auto

No injection through the loop at the 6­
way electrical valve; direct loading of
the sample on top of the column; starts
directly after equilibration (if activated)

Valve injection

Injection via the loop at the 6­way elec­
trical valve; starts directly after equili­
bration (if activated)

Wait for injection

No injection via the loop at the 6­way
electrical valve; direct loading of the
sample on top of the column which has
to be confirmed manually; starts after
equilibration

Valve inj. + Wait

Injection via the loop at the 6­way elec­
trical valve which has to be confirmed
by the user

6.6 Make a liquid injection

The loading and injection can be done without stopping the pump.

u

Fill a syringe with the volume of sample chosen and place it in the injection
adapter of the 6­way electrical valve.

• Keep the syringe on the adapter to avoid a leak of the sample.

u

Fill the loop with the content of the syringe. If the quantity injected is
greater than the capacity of the loop, the surplus flows in the waste bottle.

• Default loop volume: 2 mL

u

Create a new method to get access to the settings menu.

6.4 Create a new method [

}

page 75]

u

Select the tab Settings.