6.5 Equilibration and injection modes
u
Choose an equilibration mode to condition the column before injection.
Mode
Action
None
No column conditioning before injec
tion
Equilibration
Column conditioning before injection;
time designation
Equil. + Dry load
Column conditioning before injection;
no timing; no possibility for loop injec
tion at the 6way electrical valve
u
Choose injection mode
Mode
Action
Auto
No injection through the loop at the 6
way electrical valve; direct loading of
the sample on top of the column; starts
directly after equilibration (if activated)
Valve injection
Injection via the loop at the 6way elec
trical valve; starts directly after equili
bration (if activated)
Wait for injection
No injection via the loop at the 6way
electrical valve; direct loading of the
sample on top of the column which has
to be confirmed manually; starts after
equilibration
Valve inj. + Wait
Injection via the loop at the 6way elec
trical valve which has to be confirmed
by the user
6.6 Make a liquid injection
The loading and injection can be done without stopping the pump.
u
Fill a syringe with the volume of sample chosen and place it in the injection
adapter of the 6way electrical valve.
• Keep the syringe on the adapter to avoid a leak of the sample.
u
Fill the loop with the content of the syringe. If the quantity injected is
greater than the capacity of the loop, the surplus flows in the waste bottle.
• Default loop volume: 2 mL
u
Create a new method to get access to the settings menu.
u
Select the tab Settings.