3 electrical, connection, 1 communication and power supply, connection, Notice – BUCHI PrepChrom C-700 User Manual
Page 45
![background image](/manuals/656072/45/background.png)
BÜCHI Labortechnik AG
Installation | 5
PrepChrom C700
45/110
Followon tasks
u
5.5.3 Prep HPLC column, installation [
page 48]
5.3 Electrical, connection
5.3.1 Communication and power supply, connection
NOTICE
Risk of communication error.
RS232 cables which are not secured with the 2 screws on the plug can cause a
communication error.
u
Always tighten the 2 screws on the plug when connected.
u
Connect the electric cable RS232 (1) from COM 1 to the UV Detector
module and secure it with the 2 screws on each plug.
u
Connect the electric cable RS232 (5) from COM 2 to the injection valve
module and secure it with the 2 screws on each plug.
u
Connect the electric cable RS232 (4) from COM 3 to the pump module
and secure it with the 2 screws on each plug.
u
Connect the electric cable RS232 (3) from COM 4 to the fraction collector
and secure it with the 2 screws on each plug.
• 5.7.1 System configuration, setup [
u
Connect all 5 external power supply cables to the socket (2) of each mod
ule.