5 | Installation
BÜCHI Labortechnik AG
46/110
PrepChrom C700
5.3.2 External devices, connection
NOTICE
Risk of software damage due to malware.
No antivirus software is installed on the C700. The internal software can be
damaged by plugging in external devices.
u
The primary use of the RJ45 connection is printing.
u
The primary use of the USB connections are the use of a mouse, Key
board and memory stick.
u
Connecting to any network or connecting a memory stick requires a safe
and virus free network.
NOTE
To ensure the stability of the software it is strongly recommended not to install
any antivirus software on the control unit.
u
Connect the cable on the right side of the control unit to the RJ45 or USB
socket.
• Printer
• Keyboard
• Mouse
• USB memory stick
u
Do not install additional programs on the C700.
NOTE
The maximum length of external USB cables is limited to 2 meter.
5.4 Rack and glass tubes, installation
u
Insert the glass tubes (2) into the rack (3).
u
Place the rack in front of the guide (4).
u
Depending on the used glass tubes, check the distance to the needle (1)
of the fraction collector. The distance should be 5–25 mm.