5 | Installation
BÜCHI Labortechnik AG
68/110
PrepChrom C700
Note
Column 1 or column 2 can be also selected in the synoptic by clicking on the
column 1 or 2 (1).
5.7.15 Back flush activation/deactivation
Note
The installation of the back flush valve needs to be done by a BUCHI customer
service representative.
u
Press Setup > System configuration on the main menu and select the
tab Valves.
u
Click into the field Back Flush Valve, select Yes (1) to activate the valve,
select the valve type ACTV (2), search for the right COM port (3), deter
mine the configuration of the valve (Reverse or Normal) (4) and confirm
with OK (5).