2 flow cell, replacement, 3 flow cell windows, replacement – BUCHI PrepChrom C-700 User Manual
Page 91
![background image](/manuals/656072/91/background.png)
BÜCHI Labortechnik AG
Maintenance | 7
PrepChrom C700
91/110
u
Remove 2 countersunk screws on top of the UV Detector module.
u
Remove 4 hex screws (1) and washers.
u
Remove the cover (2).
7.3.2 Flow cell, replacement
u
Replace the flow cell according to the instruction of the flow cell cleaning.
• 7.2.2 Flow cell, cleaning [
7.3.3 Flow cell windows, replacement
Preliminary tasks
u
Remove the flow cell according the following task:
• 7.2.2 Flow cell, cleaning [
Tasks
u
Remove the 4 screws (6) and nuts (1).
u
Remove both cell covers (2).
u
Remove both window covers (3).