7 | Maintenance

BÜCHI Labortechnik AG

88/110

PrepChrom C­700

u

Purge the system with a suitable solvent or alcohol (1­propanol or 2­
propanol) during approx. 3 minutes.

• The 6­way electrical valve is on the LOAD position during a purge.

u

Click on the loop in the synoptic to switch the position of the 6­way electri­
cal valve and clean the RUN position (containing the loop or the solid de­
posit).

u

Stop the purge.

7.2.2 Flow cell, cleaning

Preliminary tasks

u

Purge the system with a suitable solvent (isopropanol).

u

Air purge the system.

• 6.9 Air purge [

}

page 85]

u

7.3.1 UV Detector cover, removal [

}

page 90]

Tasks

u

Disconnect both tubes (3) on each side of the flow cell (2).

u

Lift up the lever (1) about 45°.