6 | Operation

BÜCHI Labortechnik AG

78/110

PrepChrom C­700

Step gradient mode

Note

The first step is used as a start point only and corresponds to the condition used
during equilibration (if activated). The time of the first step can not be modified.

u

Adjust the flow rate (1).

Note

When using the BUCHI column methods, the system sets the optimal flow rate
for the column (only flash columns).

u

Adjust the percentage of the used solvents (2).

Note

Always start adjusting the solvent percentages from the right to the left.

u

Add the second step (3).

u

Set the end time for the second step (4).

Note

The flow rate and the percentage of the used solvents in the second step are
equal to the first step (= equilibration step) and can not be modified separately.

u

Add other steps for elution (3) and adjust the flow rate and percentage of
the solvents.

u

Select the tab Detection.