Metrohm 785 DMP Titrino User Manual
Page 64
2.16 Silo memory for sample data
785 DMP Titrino
61
Key
Sample data can be entered into the silo memory with
key
edit silo lines:
Entering sample data into the silo memory.
delete silo lines:
Deletes single silo lines.
delete all silo lines:
Deletes the whole silo memory.
The display texts of the Titrino are shown to the left.
The values are the default values.
>edit silo lines
silo line 1
method:
id
#1 or C21
id
#2 or C22
id
#3 or C23
smpl size 1.0 g
smpl unit: g
Input for silo memory
Silo line (1...255)
The next free line is displayed automatically. Lines
already occupied can be corrected.
Method with which the sample is processed (method
name from the method memory)
If no method name has been entered, the sample is
processed with the method in the working memory.
Selection of the method with <
←
/
→
>.
Sample identification 1..3 or sample specific calculation
variables C21...C23 (up to 8 characters)
Sample size (6-digit number: ±X.XXXXX)
The method specific limits are tested on result
calculation.
Unit of sample size (g, mg, ml, ul, pc, no unit or up to 5
characters)
Select unit with <
←
/
→
>.
>delete silo lines
delete line n OFF
Delete individual silo lines
Line number of the line to be deleted(1...255, OFF)
Deleted lines remain in the silo memory. Access is
blocked during the processing. To show that a line has
been deleted, they are marked with "*". The symbol *
indicates that the line has been deleted.
Deleted lines can be reactivated if the appropriate line
is re-edited.
SMPL
DATA
smpl data
>edit silo lines
>delete silo lines
>delete all silo lines
cycle lines: OFF
save lines: OFF