3.2. Remote control commands
785 DMP Titrino
122
Selection of the unit of which the measured values should be sent (3.2.2.123
or 3.2.2.124).
3.2.2.123. Setup.SendMeas.Assembly.CyclNo
ON, OFF
Setup.SendMeas.Assembly.V
ON, OFF
Setup.SendMeas.Assembly.Meas
ON, OFF
Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.121):
.CyclNo:
Cycle number of the potential measurement. Together with the
cycle time (3.2.2.108), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and
it is always incremented as long as the instrument remains
switched on.
.V:
Volume
.Meas:
Measured value associated to the cycle number.
The unit "assembly" must be preset (see 3.2.2.122).
3.2.2.124. Setup.SendMeas.Titrator.CyclNo
ON, OFF
Setup.SendMeas.Titrator.V
ON, OFF
Setup.SendMeas.Titrator.Meas
ON, OFF
Setup.SendMeas.Titrator.dVdt
ON, OFF
Setup.SendMeas.Titrator.dMeasdt
ON, OFF
Setup.SendMeas.Titrator.dMeasdV
ON, OFF
Setup.SendMeas.Titrator.ERC
ON, OFF
Setup.SendMeas.Titrator.T
ON, OFF
Selection of the values from the titrator which are sent in the set time interval
(see 3.2.2.121, formats see 3.2.2.104):
.CyclNo:
Cycle number. Together with the cycle time (3.2.2.108), a time
frame can be set up. The other data belong to the corresponding
cycle number. The cycle number is set to 0 at the start of a
method and it is incremented until the end of the method.
.V:
Volume.
.Meas:
Measuring value
.dVdt:
associated volume drift.
.dMeasdt: associated measured value drift.
.dMeasdV: associated 1st derivative of the titration curve.
.ERC:
ERC in DET.
.T:
Temperature.
The unit "titrator" must be preset (see 3.2.2.122).
3.2.2.125. Setup.AutoInfo.Status
ON, OFF
Setup.AutoInfo.P
ON, OFF
Setup.AutoInfo.T.R
ON, OFF
Setup.AutoInfo.T.G
ON, OFF
Setup.AutoInfo.T.GC
ON, OFF
Setup.AutoInfo.T.S
ON, OFF
Setup.AutoInfo.T.B
ON, OFF
Setup.AutoInfo.T.F
ON, OFF
Setup.AutoInfo.T.E
ON, OFF
Setup.AutoInfo.T.H
ON, OFF
Setup.AutoInfo.T.C
ON, OFF