4.1. Error and special messages

785 DMP Titrino

136

message appears each time the instrument is switched on

until it has been readjusted (Metrohm service).

system error 14

No communication between the Titrino and the connected
Remote Box.
Possible causes:
. The Remote Box was connected when the Titrino was
running
. Titrino has a fault.
. Remote Box has a fault.
Remedy: Set under , >peripheral units, "Remote
Box: OFF", switch off Titrino, take away Remote Box and
switch on Titrino. Contact Metrohm service.

time-out PC keyboard

A connected PC keyboard has been used to call up an ad-
dress (e.g. ) and the connection has then been inter-
rupted.
Possible causes:
. Remote Box has a fault.
. PC keyboard has a fault.
Exit: Correct fault and switch Titrino off/on.

TIP terminated

TIP has been terminated.

transmission error

With a Remote Box connected characters are received which
cannot be interpreted.
Possible causes:
. Wrong key combination has been pressed.
. Wrong PC keyboard has been selected.
. The barcode reader supplies garbled characters.
. The Remote Box has a fault.
Exit: Rectify fault and switch Titrino off/on.

validate instrument

Validation interval has elapsed.
Exit: or new start.

wrong card

The card was inserted/removed during an operation with

.

Exit:

wrong sample

With SET, with preset titration direction the first measured

value is outside the end point.