Metrohm 785 DMP Titrino User Manual
Page 121
3.2. Remote control commands
785 DMP Titrino
118
Info.ActualInfo.MeasPt.Y
read only
Info.ActualInfo.MeasPt.Z1
read only
Info.ActualInfo.MeasPt.Z2
read only
Info.ActualInfo.EP.Index
read only
Info.ActualInfo.EP.X
read only
Info.ActualInfo.EP.Y
read only
$Q sends the last entry into the measuring point list (.MeasPt) or the last en-
try into the list of EP's with DET, MET, SET.
.MeasPt.X"165" Volume (DET, MET), time (SET), resp. of the MPList in s
.MeasPt.Y"3.654" Measured value (DET, MET), volume (SET), resp. of the
MPList in mL
.MeasPt.Z1"6.34" Measured value (SET, MEAS) of the MPList, format de-
pends on the measured quantity
.MeasPt.Z2"25.8" Temperature of the MPList in °C
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.106. Info.ActualInfo.Display.L1
up to 32 ASCII characters
Info.ActualInfo.Display.L8
up to 32 ASCII characters
Info.ActualInfo.Display.DelAll
$G
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 3.2.2.119.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.119
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.78) to refresh the
display.
$Q sends the contents of the corresponding display line.
3.2.2.107. Info.ActualInfo.Comport.Number
read only
$Q sends the comport number of the Titrino where the PC is connected.
3.2.2.108. Info.Assembly.CycleTime
read only
Info.Assembly.ExV
read only
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.
3.2.2.109. Assembly.Bur.Rates.Forward.Selected
digital, analog
Assembly.Bur.Rates.Forward.Digital
0...150, max.
Assembly.Bur.Rates.Reverse.Selected
digital, analog
Assembly.Bur.Rates.Reverse.Digital
0...150, max.
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.