Metrohm 785 DMP Titrino User Manual

Page 122

background image

3.2 Remote control commands

785 DMP Titrino

119

3.2.2.110. Assembly.Bur.Fill

$G, $H, $C

$G starts the 'FILL' mode of the burette function.

3.2.2.111. Assembly.Bur.ModeDis

$G, $S, $H, $C

Assembly.Bur.ModeDis.Selected

volume, time

Assembly.Bur.ModeDis.V

0.0001...0.1...9999

Assembly.Bur.ModeDis.Time

0.25...1...86400

Assembly.Bur.ModeDis.VStop

0.0001...9999, OFF

Assembly.Bur.ModeDis.AutoFill

ON, OFF

Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop:

Limit volume for the dispensing.

.AutoFill:

ON means automatic filling after every dispensing.

3.2.2.112. Assembly.Meas.Status

ON, OFF

Assembly.Meas.MeasInput

1, 2, Diff., Ipol, Upol, Temp

Assembly.Meas.Ipol

±127...1...+127

Assembly.Meas.Upol

±1270...400...+1270

Measurement in assembly. The measuring function can only be started via
RS Control. When the measuring function is switched on, no method can be
started at the Titrino.
.Input:

Selection of the potentiometric measuring input 1, 2, diff., polar-
ized electrodes or temperature.

.Ipol:

Polarization current in uA.

.Upol:

Polarization potential in mV, entry in steps of 10 mV.

3.2.2.113. Assembly.Outputs.AutoEOD

ON, OFF

Assembly.Outputs.SetLines

$G

Assembly.Outputs.SetLines.L0

active, inactive, pulse, OFF

up to .L13
Assembly.Outputs.ResetLines

$G

Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the

end of the determination can be switched off. Thus, for example,
in conjunction with a Titrino several determinations can be per-
formed in the same beaker. Before AutoEOD is switched on, line 3
must be set to "OFF".

.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal,

"inactive" means resetting of the signal, "pulse" means output of a
pulse of app. 150 ms, "OFF" means the line is not operated, see
also page 166.
Warnings:

If you have "AutoEOD" to "ON", an active line 3 is set to
"inactive" by the EOD pulse.