Page 106

background image

3.2 Remote control commands

785 DMP Titrino

103

Parameters for SET: Control parameters, see page 30.
.Dyn:

Dynamics, control range in pH, mV (with U and Ipol) or uA (with
Upol). The corresponding unit can be read with .UnitDyn.

.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum

possible rate with the Exchange Unit in current use.

.MinRate: Minimum titration rate in ul/min.

3.2.2.30. Mode.Parameter.SET1.Stop.Type

drift, time

Mode.Parameter.SET1.Stop.Drift

1...20...999

Mode.Parameter.SET1.Stop.Time

0...10...999, inf

Mode.Parameter.SET1.Stop.StopT

0...99 999, OFF

Parameters for SET: Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in ul/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"

means infinite.

.StopT: Stop time in s. Applies when "time" has been selected and the value

of .Time is set to "inf.".

3.2.2.31. Mode.Parameter.TitrPara.Direction

+, -, auto

Parameters for SET: Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment. If 2 EP's have been set in a SET titration, the titration direction is given
by the two EP's. The entry of the titration direction is then invalid.

3.2.2.32. Mode.Parameter.TitrPara.XPause

0...999 999

Parameter for SET: Pause time in s. Runs before dosing the start volume.

3.2.2.33. Mode.Parameter.TitrPara.ExtrT

0...999 999

Parameter for SET: Extraction time in s.

3.2.2.34. Mode.Parameter.TitrPara.TDelta

1...2...999 999

Parameter for SET: Time interval in s for the entry of a measurement point in
the list of measured points.

3.2.2.35. Mode.Parameter.Presel.Cond

ON, OFF

Mode.Parameter.Presel.DriftDisp

ON, OFF

Mode.Parameter.Presel.DCor.Type

auto, man., OFF

Mode.Parameter.Presel.DCor.Value

0.0...99.9

Parameters for SET:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of

the drift value at start.

.DCor.Value: Drift value for the manual drift correction.