3.2 Remote control commands

785 DMP Titrino

111

3.2.2.81. Config.RSSet1

$G

Config.RSSet1.Baud

300, 600, 1200, 2400, 4800, 9600,

19200, 38400, 57600, 115200

Config.RSSet1.DataBit

7, 8

Config.RSSet1.StopBit

1, 2

Config.RSSet1.Parity

even, odd, none

Config.RSSet1.Handsh

HWs, SWchar, SWline, none

$G sets all RS settings. The changes are performed only if the instrument is
inactive. After the setting of the interface parameters, wait at least 2 s to allow
the components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud
rate, data bit, stop bit, parity and type of handshake, see also page 126ff.
Baud rates >9600 need a PC which is equipped accordingly (e.g. with
16550 component).
The setting of the values must be initiated with $G immediately after entry of
the values.

3.2.2.82. Config.ComVar.C30

with up to .C39, etc.

0... +999 999

Values of the common variables from C30 up to C39. Insert the common
variables directly or describe the determination results directly from the
method, see 3.2.2.7

3.2.2.83. SmplData.Status

ON, OFF

On/off switching of silo memory. When the silo memory is switched on, the
sample data are fetched from the lowest valid silo line.

3.2.2.84. SmplData.OFFSilo.Id1

up to 8 ASCII characters

SmplData.OFFSilo.Id2

up to 8 ASCII characters

SmplData.OFFSilo.Id3

up to 8 ASCII characters

SmplData.OFFSilo.ValSmpl

6-digits, sign and decimal point

SmplData.OFFSilo.UnitSmpl

up to 5 ASCII characters

Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific cal-
culation constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be
entered.

3.2.2.85. SmplData.ONSilo.Counter.MaxLines

read only

SmplData.ONSilo.Counter.FirstLine

read only

SmplData.ONSilo.Counter.LastLine

read only

Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.

3.2.2.86. SmplData.ONSilo.EditLine.1.Method

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id1

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id2

up to 8 ASCII characters