3.2. Remote control commands

785 DMP Titrino

110

.Keyboard: Type of keyboard which is connected to the Remote Box.
.Barcode: Select target in Titrino where you wish to have the string from the

barcode reader. "input" means that the string comes into the field
where the cursor is currently placed.

3.2.2.73. Config.Aux.Language

english, deutsch, francais, espanol,

italiano, portugese, svenska

Selection of the dialog language.

3.2.2.74. Config.Aux.Set

$G

Config.Aux.Set.Date

YYYY-MM-DD

Config.Aux.Set.Time

HH:MM

Date and time.
Input format of the date: Year-month-day, two-digit, enter leading zeros.
Input format for the time: Hours:minutes, two-digit, enter leading zeros.
Date and time have to be set with &Config.Aux.Set $G just after entry of the
value.

3.2.2.75. Config.Aux.RunNo

0...9999

Current sample number.
Set to 0 on power on and initialization. After 9999, counting starts again at 0.

3.2.2.76. Config.Aux.AutoStart

1...9999, OFF

Number of automatic, internal starts.

3.2.2.77. Config.Aux.StartDelay

0...999 999

Start delay time in s. During this time, the data of the preceding determination
are retained.

3.2.2.78. Config.Aux.ResDisplay

bold, standard

Character set for the result display at the end of the determination.

3.2.2.79. Config.Aux.DevName

up to 8 ASCII characters

Name of the instrument for connections with several units. It is advisable to
use only the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and
the numbers 0...9 (ASCII No. 48...57) when the function Setup.AutoInfo
(3.2.2.125) is used at the same time.
If a name has been entered, it will be printed out in the result report (full,
short).

3.2.2.80. Config.Aux.Prog

read only

Output of the program version.
The Titrino sends "785.0010" on requests with $Q.