beautypg.com

Metrohm 785 DMP Titrino User Manual

Page 118

background image

3.2 Remote control commands

785 DMP Titrino

115

Measured value coordinate in pH "5.12", mV (with U and Ipol) "-
241" or uA (with Upol) "43.7".

.Var: Various variables. You may overwrite the variables C40...C45, see

3.2.2.95.
C40: Initial measured value in pH "5.12", mV (with U and Ipol)
"41", uA (with Upol) "43.7" or °C (with T) "25.0". In MEAS final
measured value.
C41: End volume with SET in ml, "12.5360".
C42: Time from start of titration to end in s, "62".
C43: Volume drift on start of a SET titration from the conditioning
in ul/min, "3.5".
C44: (Last measured) temperature in °C. Used for the tempera-
ture compensation in pH measurements.
C45: Start volume with DET, MET, SET in ml, "2.800".
C46: Asymmetry pH of CAL, "6.89".
C47: Relative electrode slope of CAL, "0.9950".
C48: Voltage at maximum voltage (not valid in CAL and TIP)
C49: Voltage at minimum voltage (not valid in CAL and TIP)
DTime: Time in s for the drift correction in SET with conditioning.

3.2.2.97. Info.TitrResults.FixEP.51.Value

read only

etc. up to .59
Info.TitrResults.pK.61.Value

read only

etc. up to .69
Info.TitrResults.TempVar.C70

read only/read+write

etc. up to .C79

.FixEP: Fix EP with DET, MET. C5X corresponds to X = 1...9.
.pK: With DET, MET. C6X corresponds to X = 1...9.
.TempVar: Temporary variables in TIP corresponding to the assignments in

the submethods.

3.2.2.98. Info.StatisticsVal.ActN

read only

Info.Statistics.1.Mean

read only

Info.Statistics.1.Std

read only

Info.Statistics.1.RelStd

read only

etc. up to .9

The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results

"3"

Data for MN1:
Mean: Mean value (decimal places as in result)

"3.421"

Std: Standard deviation (1 decimal place more than in result)

"0.0231"

RelStd: Relative standard deviation (in %, 2 decimal places)

"0.14"

3.2.2.99. Info.SiloCalc.C24.Name

read only

Info.SiloCalc.C24.Value

read only

Info.SiloCalc.C24.Unit

read only

for .C25 as for .C24
Info.SiloCalc.C26.ActN

read only

Info.SiloCalc.C26.Mean

read only