Metrohm 785 DMP Titrino User Manual

Page 112

background image

3.2 Remote control commands

785 DMP Titrino

109

3.2.2.65. Config.Monitoring.Calibration.Status

ON, OFF

Config.Monitoring.Calibration.MeasInput

1, 2, diff

Config.Monitoring.Calibration.Interval

1...7...9999

Config.Monitoring.Calibration.Counter

0...9999

Monitoring of pH calibration.
.MeasInput:Measuring input.
.Interval:

Time interval in days for calibration from the last calibration date.

.Counter: Time counter in days since last calibration. The counter is reset to

zero if a new calibration is carried out or if the calibration for the
corresponding measuring input is entered manually.

3.2.2.66. Config.Monitoring.Service.Status

ON, OFF

Config.Monitoring.Service.Date

XXXX-XX-XX

Monitoring of service interval.

3.2.2.67. Config.Monitoring.DiagRep

ON, OFF

Printing of system test report after each switching on of the Titrino.

3.2.2.68. Config.PeriphUnit.CharSet1

Epson, Seiko, Citizen, HP, IBM

Config.PeriphUnit.CharSet2

Selection of the character set and the graphics control characters for COM1
resp. COM2 of the Titrino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.

3.2.2.69. Config.PeriphUnit.RepToComport

1, 2, 1&2

Selection of COM of the Titrino where manually triggered reports should be
outputted.

3.2.2.70. Config.PeriphUnit.Balance

Sartorius,Mettler,Mettler AT,

AND,Precisa

Selection of the balance type.

3.2.2.71. Config.PeriphUnit.Stirrer

ON, OFF

Automatic stirrer control. With "ON" the stirrer will be switched on after start-
ing a method. At the end of the method it is switched off again. Valid for DET,
MET, SET.

3.2.2.72. Config.PeriphUnit.RemoteBox.Status

ON, OFF

Config.PeriphUnit.RemoteBox.Keyboard

US, deutsch, francais,

español, schweiz.

Config.PeriphUnit.RemoteBox.Barcode

input, method, id1, id2,

id3, smpl size

Connections via Remote Box.
.Status:

Select if a Remote Box is connected.