Metrohm 785 DMP Titrino User Manual
Page 112
3.2 Remote control commands
785 DMP Titrino
109
3.2.2.65. Config.Monitoring.Calibration.Status
ON, OFF
Config.Monitoring.Calibration.MeasInput
1, 2, diff
Config.Monitoring.Calibration.Interval
1...7...9999
Config.Monitoring.Calibration.Counter
0...9999
Monitoring of pH calibration.
.MeasInput:Measuring input.
.Interval:
Time interval in days for calibration from the last calibration date.
.Counter: Time counter in days since last calibration. The counter is reset to
zero if a new calibration is carried out or if the calibration for the
corresponding measuring input is entered manually.
3.2.2.66. Config.Monitoring.Service.Status
ON, OFF
Config.Monitoring.Service.Date
XXXX-XX-XX
Monitoring of service interval.
3.2.2.67. Config.Monitoring.DiagRep
ON, OFF
Printing of system test report after each switching on of the Titrino.
3.2.2.68. Config.PeriphUnit.CharSet1
Epson, Seiko, Citizen, HP, IBM
Config.PeriphUnit.CharSet2
Selection of the character set and the graphics control characters for COM1
resp. COM2 of the Titrino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.69. Config.PeriphUnit.RepToComport
1, 2, 1&2
Selection of COM of the Titrino where manually triggered reports should be
outputted.
3.2.2.70. Config.PeriphUnit.Balance
Sartorius,Mettler,Mettler AT,
AND,Precisa
Selection of the balance type.
3.2.2.71. Config.PeriphUnit.Stirrer
ON, OFF
Automatic stirrer control. With "ON" the stirrer will be switched on after start-
ing a method. At the end of the method it is switched off again. Valid for DET,
MET, SET.
3.2.2.72. Config.PeriphUnit.RemoteBox.Status
ON, OFF
Config.PeriphUnit.RemoteBox.Keyboard
US, deutsch, francais,
español, schweiz.
Config.PeriphUnit.RemoteBox.Barcode
input, method, id1, id2,
id3, smpl size
Connections via Remote Box.
.Status:
Select if a Remote Box is connected.