785 DMP Titrino

99

Mode.CFmla.1.Value

0...

±

999 999

Mode.CFmla.2.Value
etc., up to .19

Calculation constants specific to a method. Stored in the method memory of
the Titrino. Operands specific to the sample (3.2.2.84 and 3.2.2.85) and val-
ues of common variables (3.2.2.82) on the other hand are not stored with the
methods.

3.2.2.12. Mode.Parameter.TitrPara.MptDensity

0...4...9

Mode.Parameter.TitrPara.MinIncr

0...10.0...999.9

Mode.Parameter.TitrPara.VStep

0...0.10...9.999

.MptDensity: Parameter for DET: Measuring point density.
.MinIncr: Parameter for DET: Minimum increment in uL. If the minimum in-

crement is set to 0, measured values are stored vs. time.

.VStep:

Parameter for MET: Volume increment in mL. With "0", there is no
dispensing and measured values vs. time are entered in the
measuring point list.

3.2.2.13. Mode

.

Parameter.TitrPara.DosRate

0.01...150, max.

Parameters for DET and MET: Dispensing rate for the volume increments in
mL/min. Max. means maximum possible dispensing rate with the Exchange
Unit in current use.

3.2.2.14. Mode.Parameter.TitrPara.SignalDrift pH,U,Ipol:0.5...50...999, OFF

Upol: 0.05...50...99.9, OFF

Mode.Parameter.TitrPara.UnitSigDrift

read only

Mode.Parameter.TitrPara.EquTime

0...26...9999, OFF

Parameters for DET and MET: Criteria for the measured value acquisition.
Measured value drift in mV/min (with pH, U, Ipol) or uA/min (with Upol),
equilibration time in s. OFF means that the corresponding criterion is switched
off. If both criteria are OFF, the measured values are acquired immediately
after dispensing.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 19. After it has been edited once, it
remains in force with the set value.

3.2.2.15. Mode.Parameter.TitrPara.StartV.Type

abs., rel., OFF

Mode.Parameter.TitrPara.StartV.V

0...999.99

Mode.Parameter.TitrPara.StartV.Factor

0...±999 999

Mode.Parameter.TitrPara.StartV.Rate

0.01...150, max.

Parameters for DET, MET, SET: Start volume.
If an absolute start volume (abs.) has been selected, the volume in mL is
valid.
A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size

*

factor

The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.