3.2. Remote control commands

785 DMP Titrino

102

.Recognition.Selected: EP recognition.

all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.

.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U

and Ipol) or uA (with Upol).

.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and

Ipol) or uA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 22.

.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. uA (for Upol). Fix EP's are

evaluated until the setting OFF is found.

.pK: pK or HNP evaluation. Possible only in pH and U titrations.

3.2.2.25. Mode.Parameter.Presel.IReq

id1, id1&2, all, OFF

Mode.Parameter.Presel.SReq

value, unit, all, OFF

Parameters for DET, MET, SET, MEAS: Automatic inquiry after the start of the
determination. From such an inquiry, the determination continues if the re-
quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.84)
or with &M $G, see 3.2.2.1.
$H is not possible in requests.

3.2.2.26. Mode.Parameter.Presel.LimSmplSize.Status

ON, OFF

Mode.Parameter.Presel.LimSmplSize.LoLim

0.0...999 999

Mode.Parameter.Presel.LimSmplSize.UpLim

0.0...999 999

Limit control for the sample size.

3.2.2.27. Mode.Parameter.Presel.ActPuls

ON, OFF

for SET:

first, all, cond., OFF

Output of a pulse on the I/O line "Activate", see page 166.

3.2.2.28. Mode.Parameter.SET1.EP

pH:

0...

±

20.00, OFF

U, Ipol: 0...

±

2000, OFF

Upol:

0...

±

200.0, OFF

Mode.Parameter.SET1.UnitEp

read only

Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. uA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.

3.2.2.29. Mode.Parameter.SET1.Dyn

pH: 0.01...20.00, OFF
U, Ipol: 1...2000, OFF
Upol: 0.1...200.0, OFF

Mode.Parameter.SET1.UnitDyn

read only

Mode.Parameter.SET1.MaxRate

0.01...10...150, max.

Mode.Parameter.SET1.MinRate

0.01...25.0...999.9