3.2. Remote control commands
785 DMP Titrino
102
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.
.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U
and Ipol) or uA (with Upol).
.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and
Ipol) or uA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 22.
.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. uA (for Upol). Fix EP's are
evaluated until the setting OFF is found.
.pK: pK or HNP evaluation. Possible only in pH and U titrations.
3.2.2.25. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Parameters for DET, MET, SET, MEAS: Automatic inquiry after the start of the
determination. From such an inquiry, the determination continues if the re-
quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.84)
or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.26. Mode.Parameter.Presel.LimSmplSize.Status
ON, OFF
Mode.Parameter.Presel.LimSmplSize.LoLim
0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim
0.0...999 999
Limit control for the sample size.
3.2.2.27. Mode.Parameter.Presel.ActPuls
ON, OFF
for SET:
first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 166.
3.2.2.28. Mode.Parameter.SET1.EP
pH:
0...
±
20.00, OFF
U, Ipol: 0...
±
2000, OFF
Upol:
0...
±
200.0, OFF
Mode.Parameter.SET1.UnitEp
read only
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. uA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.29. Mode.Parameter.SET1.Dyn
pH: 0.01...20.00, OFF
U, Ipol: 1...2000, OFF
Upol: 0.1...200.0, OFF
Mode.Parameter.SET1.UnitDyn
read only
Mode.Parameter.SET1.MaxRate
0.01...10...150, max.
Mode.Parameter.SET1.MinRate
0.01...25.0...999.9