beautypg.com

Metrohm 785 DMP Titrino User Manual

Page 11

background image

2.4. Configuration, key

785 DMP Titrino

8

service: OFF

next service YYYY-MM-DD

system test report: OFF

Monitoring the service interval (ON, OFF)

Monitoring is carried out after the Titrino has been

switched on. If the monitoring responds the message

"Service is due" appears. The message vanishes with

.

If "on" has been set:

Date of next service (YYYY-MM-DD)

System test report printout (ON, OFF)

With "on" the report of the system test is printed out

after the Titrino has been switched on, see also page

168.

>peripheral units

send to COM1: IBM

send to COM2: IBM

man.reports to COM: 1

balance: Sartorius

stirrer control: OFF

Settings for peripheral units

Selection of printer (Epson, Seiko, Citizen, HP, IBM) at

the Titrino COM1

"Epson", for Epson

"Seiko", e.g. for DPU-414

"Citizen", e.g. for iDP 562 RS

"HP" e.g. for Desk Jet types. Place curves always at the

beginning of a page as you cannot have them over 2

pages.

"IBM" for all printers with IBM character set Table 437

and IBM graphics, as well as for the data transmission

to a computer or a data system.

COM of Titrino for the output of manually triggered

reports (1, 2, 1&2)

Manually triggered reports e.g. with .... .

Exception : These reports are

outputted on the COM as defined in the method.

Selection of balance (Sartorius, Mettler, Mettler AT,

AND, Precisa)

Sartorius:

Models MP8, MC1

Mettler:

Models AM, PM and balances with 011,

012, and 016 interfaces

Mettler AT: Model AT

AND:

Models ER-60, 120, 180, 182, FR-200, 300

and FX-200, 300, 320

Precisa:

Models with RS232C interface

Automatic switching ON/OFF of the stirrer in the titration

sequence (ON, OFF)

If stirrer control is ON, the stirrer will be switched on at
the beginning and switched OFF at the end of a de-
termination. For SET with conditioning the stirrer will be
switched off in the inactive state.
In the modes MEAS, CAL, and TIP the stirrer is not
switched automatically.
For stirrer control the red switch on the stirrer unit must
be ON.