Metrohm 785 DMP Titrino User Manual
Page 11
2.4. Configuration, key
785 DMP Titrino
8
service: OFF
next service YYYY-MM-DD
system test report: OFF
Monitoring the service interval (ON, OFF)
Monitoring is carried out after the Titrino has been
switched on. If the monitoring responds the message
"Service is due" appears. The message vanishes with
If "on" has been set:
Date of next service (YYYY-MM-DD)
System test report printout (ON, OFF)
With "on" the report of the system test is printed out
after the Titrino has been switched on, see also page
168.
>peripheral units
send to COM1: IBM
send to COM2: IBM
man.reports to COM: 1
balance: Sartorius
stirrer control: OFF
Settings for peripheral units
Selection of printer (Epson, Seiko, Citizen, HP, IBM) at
the Titrino COM1
"Epson", for Epson
"Seiko", e.g. for DPU-414
"Citizen", e.g. for iDP 562 RS
"HP" e.g. for Desk Jet types. Place curves always at the
beginning of a page as you cannot have them over 2
pages.
"IBM" for all printers with IBM character set Table 437
and IBM graphics, as well as for the data transmission
to a computer or a data system.
COM of Titrino for the output of manually triggered
reports (1, 2, 1&2)
Manually triggered reports e.g. with
Exception
outputted on the COM as defined in the method.
Selection of balance (Sartorius, Mettler, Mettler AT,
AND, Precisa)
Sartorius:
Models MP8, MC1
Mettler:
Models AM, PM and balances with 011,
012, and 016 interfaces
Mettler AT: Model AT
AND:
Models ER-60, 120, 180, 182, FR-200, 300
and FX-200, 300, 320
Precisa:
Models with RS232C interface
Automatic switching ON/OFF of the stirrer in the titration
sequence (ON, OFF)
If stirrer control is ON, the stirrer will be switched on at
the beginning and switched OFF at the end of a de-
termination. For SET with conditioning the stirrer will be
switched off in the inactive state.
In the modes MEAS, CAL, and TIP the stirrer is not
switched automatically.
For stirrer control the red switch on the stirrer unit must
be ON.