Page 104

background image

3.2 Remote control commands

785 DMP Titrino

101

3.2.2.21. Mode.Parameter.StopCond.EPStop

1...9, OFF

Parameters for DET and MET: Stop when a certain number of EP's has been
found.
OFF means that the criterion is not monitored.

3.2.2.22. Mode.Parameter.StopCond.FillRate

0.01...150, max.

Parameters for DET, MET, SET: Filling rate in the titration in mL/min. Max.
means maximum possible filling rate with the Exchange Unit in current use.

3.2.2.23. Mode.Parameter.Statistics.Status

ON, OFF

Mode.Parameter.Statistics.MeanN

2...20

Mode.Parameter.Statistics.ResTab.Selected

original, delete n,

delete all

Mode.Parameter.Statistics.ResTab.DelN

1...20

Entries for the statistics calculations.
.Status:

On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.

.MeanN:

Number of individual results for statistics calculations.

.ResTab.Select: Selection of the table for the statistics calculations.

original: Original table. The original table is (again) set up, i.e. any

individual results which have been deleted are reincorporated
in the statistics calculations.

delete n: Single result lines are removed from the statistics calcu-

lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.

delete all: Clear entire statistics table. The results can not be reac-

tivated.

.ResTab.DelN: Specification of the line number to be deleted.

3.2.2.24. Mode.Parameter.Evaluation.EPC

DET: 0...5...200
MET pH: 0.1...0.50...9.99

U, Ipol: 1...30...999
Upol:

0.1...2...99.9

Mode.Parameter.Evaluation.Recognition.Selected

all, greatest,

last, window, OFF

Mode.Parameter.Evaluation.Recognition.Window.1.LowLim

pH:

0...

±

20.00, OFF

U, Ipol:

0...

±

2000, OFF

Upol:

0...

±

200.0, OFF

Mode.Parameter.Evaluation.Recognition.Window.1.UpLim

Input range as LowLim

etc. up to 9 windows
Mode.Parameter.Evaluation.FixEP.1.Value pH:

0...

±

20.00, OFF

U; Ipol: 0...

±

2000, OFF

etc. up to 9 fix EP's

Upol: 0...

±

200.0, OFF

Mode.Parameter.Evaluation.pK

ON, OFF

Parameters for DET and MET: Evaluation of the EP's, see page 20.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in uA (with Upol).