3.2 Remote control commands

785 DMP Titrino

117

The cycle number is set to 0 on switching on the instrument, on every start
and for QuickMeas. It is incremented as long as the instrument remains
switched on.

3.2.2.102. Info.ActualInfo.Assembly.Counter.V

read only

Info.ActualInfo.Assembly.Counter.Clear

$G

$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G,
the volume counter is set to zero.

3.2.2.103. Info.ActualInfo.Assembly.Meas

read only

$Q sends the current measured value from the assembly.

3.2.2.104. Info.ActualInfo.Titrator.CyclNo

read only

Info.ActualInfo.Titrator.V

read only

Info.ActualInfo.Titrator.Meas

read only

Info.ActualInfo.Titrator.dVdt

read only

Info.ActualInfo.Titrator.dMeasdt

read only

Info.ActualInfo.Titrator.dMeasdV

read only

Info.ActualInfo.Titrator.ERC

read only

Info.ActualInfo.Titrator.T

read only

$Q sends the current values in the following formats:

DET

MET

SET

MEAS

CAL

CyclNo

127

127

127

127

127

V(ml)

1.2345

1.2345

1.2345

-

-

Meas:

pH
U, Ipol (mV)
Upol (uA)
T (°C)

3.345
-345.6
-12.5
-

3.345
-345.6
-12.5
-

3.6(mV)
-345.6
-12.5
-

3.345
-345.6
-12.5
25.0

3.345
-
-
-

dVdt (ul/s)

-

-

2.5142

-

-

dMeasdt

pH,U,Ipol mV/s
Upol (uA/s)
T (°C/s)

0.7957
0.7957
-

0.7957
0.7957
-

0.7957
0.7957
-

0.7957
0.7957
0.7957

0.7957
-
-

dMeasdV (mV/ul) -

-

10.6326 -

-

ERC

34

-

-

-

-

T (°C)

25.9

25.9

-

25.9

-

NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal
drift is NV.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.108). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.

3.2.2.105. Info.ActualInfo.MeasPt.Index

read only

Info.ActualInfo.MeasPt.X

read only