Page 124

background image

3.2 Remote control commands

785 DMP Titrino

121

The Titrino automatically reports when a value has been confirmed with
at the Titrino. Message, e.g.:

&SmplData.OFFSilo.Id1"Trace"

The beginning of the message is marked by a space (ASCII 32).

3.2.2.119. Setup.Lock.Keyboard

ON, OFF

Setup.Lock.Config

ON, OFF

Setup.Lock.Parameter

ON, OFF

Setup.Lock.SmplData

ON, OFF

Setup.Lock.UserMeth.Recall

ON, OFF

Setup.Lock.UserMeth.Store

ON, OFF

Setup.Lock.UserMeth.Delete

ON, OFF

Setup.Lock.Display

ON, OFF

ON means disable the corresponding function:
.Keyboard: Disable all keys of the Titrinos
.Config:

Disable the key

.Parameter:Disable the key
.SmplData: Disable the key
.UserMeth.Recall: Disable "recall" in key
.UserMeth.Store: Disable "store" in key
.UserMeth.Delete: Disable "delete" in key
.Display: Disable the display, i.e. it will not be written to by the device pro-

gram of the Titrino and can be operated from the computer.

3.2.2.120. Setup.Mode.StartWait

ON, OFF

Setup.Mode.FinWait

ON, OFF

Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method or submethod in TIP

(holding point after AutoInfo !".T.GC").

.FinWait:

Holding point at the end a method or submethod in TIP (holding
point after AutoInfo !".T.F").

3.2.2.121. Setup.SendMeas.SendStatus

ON, OFF

Setup.SendMeas.Interval

0.08...4...16200, MPList

.SendStatus: ON means the automatic transmission of measured values (see

3.2.2.123 and 3.2.2.124) in the inputted interval is active.

.Interval:

Time interval (in s) for the automatic transmission of associated
measured values defined under points 3.2.2.123 and 3.2.2.124.
The inputted value is rounded off to a multiple of 0.08. The small-
est possible time interval depends on the number of measured
values which have to be sent, on the baud rate, on the load on the
interface and on the type of device connection. With "MPList" the
measured values are sent at the time of their entry into the meas-
ured point list.

The automatic transmission is switched on/off with 'SendStatus'.

3.2.2.122. Setup.SendMeas.Select

Assembly, Titrator