Metrohm 785 DMP Titrino User Manual

Page 20

background image

2.6 Parameters, key

785 DMP Titrino

17

stop EP 9
**titr.

filling rate max. ml/min
**titr.

Stop after a number of EP's have been found (1...9,

OFF)

sets "OFF".

"OFF" means that the criterion is not monitored.

Filling rate (0.01...150 mL/min, max.)

sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

>evaluation

EPC 5

EP recognition: all

low lim.1 pH -20.00

up lim.1 pH 20.00

EP evaluation/recognition

See page 20 ff.

Equivalence point criterion (input range depends on

mode.

For DET:

0...200

For MET:

pH:

0.10...9.99

U, Ipol:

1...999 mV

Upol:

0.1...99.9 uA)

Threshold for the size of the jump, see page 22.

Recognition of EP's which fulfill the EP criterion. (all,

greatest, last, window, OFF)

Selection of equivalence point recognition:

"all":

All equivalence points are recognized.

"greatest": Only the greatest (steepest) equivalence

point is recognized.

"last":

Only the last equivalence point is

recognized.

"window": Only EP's in specified windows are

recognized.

"OFF":

Evaluation switched off.

If "window" is selected, lower and upper limits of

windows are inquired.

Lower limit of window 1

and

upper limit of window 1

(Input ranges for both inquiries depend on the

measured quantity):

pH:

0...±20.00, OFF

U, Ipol:

0...±2000 mV, OFF

Upol:

0...±200.0 uA, OFF)

sets "OFF".

Only equivalence points are recognized which lie within

the set lower and upper limits. The equivalence point

numbering is defined with the windows, see page 22.

Window inquiries continue until the lower limit is set to

"OFF". Up to 9 possible windows.

Always set both limits to

OFF for a valid window.