4.1. Error and special messages

785 DMP Titrino

134

check T-sensor

No temperature sensor is attached (with MEAS T or if the

temperature monitoring is active).

Exit: Connect Pt100 or Pt1000 or .

DX overload

The motor of dosing drive X has reached ist current limit.

Exit: then . The dosing drive will be

initialized. If the error reappears after , check the

wet part: Can the cock and/or piston be moved?

data lost

The data of the memory card are lost as the battery tension

was lower than 2.37 V.

Exit: .

data set reevaluation

Message if Titrino is reevaluating downloaded measuring point

lists.

directory already exists

The directory exists already.

Remedy: and enter a different directory name or

delete the directory with the same name.

division by zero

The result could not be calculated as a divisor in the formula

was equal to zero.

Exit: Enter appropriate value.

EP overflow

In a DET or MET titration 9 EP's or more were found. The first 9

EP's will be listed.

Remedy: Recalculation of data with higher EP criterion.

manual stop

The determination has been manually stopped.

meas.pt list overflow

Maximum 500 measured points can be stored.

Exit: Use start criteria or select larger time interval.

missing EP

An EP needed for calculation in a formula is missing.

missing fix EP

A fix EP needed for calculation in a formula is missing.

no.EP not corresponding

In DET or MET titrations, the number of EP's actually found

does not match the set windows: Exactly 1 EP per window has

not been found.

no EP set

In SET, no EP has been set.

Exit: and set EP.

no meas.quantity

In TIP a manual measurement () has been

performed without defining a measuring quantity.

Exit: and define measured quantity.

no memory card

The memory card is not inserted (correctly).

Remedy: or rectify fault.

no method

The method required by the sample data from the silo memory

or in a TIP sequence is not available in the method memory.

Exit: .

no new com.var.

The common variable could not be assigned as the result or

the mean value could not be calculated. The old value remains

in force.

no new mean

No new mean value has been calculated as at least one

quantity stipulated for mean value calculations could not be

calculated.

no new silo result

No new silo result C24 or C25 could be stored as the assigned

quantity could not be calculated.

no new temp.var.

No new temporary variable C7X could be stored as the

assigned quantity could not be calculated.