beautypg.com

Metrohm 785 DMP Titrino User Manual

Page 108

background image

3.2 Remote control commands

785 DMP Titrino

105

3.2.2.42. Mode.Parameter.Calibration.Buffer.1.Value

0...7.00...

±

20.00

Mode.Parameter.Calibration.Buffer.2.Value 0...4.00...

±

20.00, OFF

etc. up to 9 buffers

Parameters for CAL: pH of buffers. The first buffer which is set to "OFF" de-
termines the number of buffers in the calibration.

3.2.2.43. Mode.Parameter.Calibration.SignalDrift

0.5...2...999, OFF

Mode.Parameter.Calibration.EquTime

0...110...9999, OFF

Parameters for CAL: Criteria for measured value acquisition. Measured value
drift in mV/min, equilibration time in s. OFF means that the corresponding
criterion is switched off. If both criterions are on OFF, the measured value is
acquired immediately.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 19. After it has been edited once, it
remains in force with the set value.

3.2.2.44. Mode.Parameter.Calibration.ElectrodeId

up to 8 ASCII char

Parameters for CAL: Electrode identification. It is classified under calibration
data, see 3.2.2.93.

3.2.2.45. Mode.Parameter.Calibration.SmplChanger

ON, OFF

Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.

3.2.2.46. Mode.Parameter.Calibration.ActPulse

first, all, OFF

Parameters for CAL: Output of a pulse on the I/O line "Activate", see page
167.

3.2.2.47. Mode.Parameter.Sequence.X.Select

method, pause, L4 output,

L6 output, info, stirrer, OFF

Parameters for TIP: Selection of an element for step X (X = 1...30). For the
parameters of the elements see 3.2.2.48.

3.2.2.48. Mode.Parameter.Sequence.X.Method

Method name

Mode.Parameter.Sequence.X.Pause

0...999 999, INF

Mode.Parameter.Sequence.X.L4Output

active, inactive,

pulse, OFF

Mode.Parameter.Sequence.X.L6Output

as for L4

Mode.Parameter.Sequence.X.Info

up to 16 ASCII characters

Mode.Parameter.Sequence.X.Stirrer

ON, OFF

Parameters for TIP: Parameters of the elements of TIP.
.Method: Method name of a method available in the user memory or on the

card. Up to 8 ASCII characters.

.Pause:

Pause time in s. INF means infinite. Continue the sequence with
&m $G.