beautypg.com

Metrohm 785 DMP Titrino User Manual

Page 127

background image

3.2. Remote control commands

785 DMP Titrino

124

3.2.2.79). Special ASCII characters in the device name are ignored. If no de-
vice name has been entered, only ! is sent. Finally the Titrino sends the infor-
mation which node has triggered the message.
Example: !John".T.Si": The message was triggered from instrument "John",
node .T.Si

3.2.2.126. Setup.Graphics.Grid

ON, OFF

Setup.Graphics.Frame

ON, OFF

Setup.Graphics.Scale

Full, Auto

Setup.Graphics.Recorder.Right

0.2...0.5...1.00

Setup.Graphics.Recorder.Feed

0.01...0.05...1.00

Change in the appearance and the format of the curve for the output. The set-
tings are valid for both Titrino COM ports.
.Grid:

On/off switching of grid over curve.

.Frame:

On/off switching of frame surrounding the curve. If grid and frame
are switched off, the curve is printed faster as the print head does
not have to move to the end of the paper.

.Scale:

Type of scaling of the measured value axis: Full means that the
scale runs from the smallest up to the greatest measured point.
With auto, the smallest measured value is taken and the next
smaller tick defines the beginning of the scale; the next greater
tick to the greatest measured value is the end of the scale.

.Right:

Relative specification of the width of the output medium (e.g. pa-
per width) for the length of the measured value axis. 1 means the
measured value axis is plotted over the entire width of the paper
(largest possible width). In extreme cases, the writing of the right
tick may lie outside.

.Feed:

Length of the volume axis referred to the burette cylinder volume,
V(B) per cm (0.1 means, e.g. 1 mL/cm with a 10 mL Exchange
Unit). Depending on the printer, the measure in cm may not al-
ways be correct.

3.2.2.127. Setup.PowerOn

$G

Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.

3.2.2.128. Setup.Initialise

$G

Setup.Initialise.Select

ActMeth, Silo, Calib, Config,

Assembly, Setup, All

Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for

the data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Calib:

pH calibration data for all measuring inputs.

Config:

All values under &Config.

Assembly: All values under &Assembly.
Setup:

All values under &Setup.

All:

Values of the entire tree (except silo and method memory).

The action must be triggered with &Setup.Initalise $G.