4.2.

Diagnosis

785 DMP Titrino

145

• Press <5>.

RS232 test 1...2

8.1 RS232 test 1

• Press <1>.

RS232 test-connector? 1

• Insert the 3.496.8560 test plug in port ‘A1’.
• Press .

Test

sequence:

1. In case of an error an error message is displayed. If for example no test plug is

connected, error message

error 68

appears).

2. If no error is found, after about 5 s display shows

RS232 test 1 o.k.

• Remove test plug.
• Press .

RS232 test 1...2

8.2 RS232 test 2

• Press <2>.

RS232 test-connector? 2

• Insert the 3.496.8560 test plug in port ‘A2’.
• Press .

Test

sequence:

1. In case of an error an error message is displayed. If for example no test plug is

connected, error message

error 68

appears)

2. If no error is found, after about 5 s display shows

RS232 test 2 o.k.

• Remove test plug.
• Press .

RS232 test 1...2

• Press .

diagnose press key 0...9